Creep testing machine

Creep testing machine is mainly usedfor testingthemechanical properties of superalloys, titanium alloys, other ferrous metals and non-ferrous metals, andmeasuring the steady-state creep rate, creep limit, creep elongation, Permanent fracture time, permanent strength limit, permanent elongation after fracture, section shrinkage, permanent notch sensitivity coefficient.Inaddition to conventional creep and durability tests, stress relaxation tests, high-temperature short-timetensiletests, cyclic loading tests (load spectrum tests), and low-cycle cycles can also be performed.

Compression, Stress Relaxation Tests, High-temperature Short-Time Tensile Tests, Cyclic Loading Tests (Load Spectrum Tests) And Low-Cycle Cycles Testing Machine

Creep test machine is applicable to do creep and durable strength test in the condition of high temperatureforallkinds of metal and alloys materials, to test creep limit, durable strength limit etc. of material.

This creep test machine is applicable to the field of quality measurement, aerospace, iron and steel metallurgy,machinery manufacturing, electronic appliances, automobile production, civilian nuclear energy, civil aviation, universities, research institutes, commodity inspection, arbitration

AASTM E1012《Standard Practice for Verification of Testing Frame and Specimen

Alignment Under Tensile and Compressive Axial Force Application1》

JB/T9373-1999 tensile creep test machine technology specifications

JJG276 high temperature creep and durable test machine

ISO 204 Metallic materials uniaxial creep testing in tension.

ISO/R 206 Creep stress rupture testing of steel at elevated temperature

creep testing machine

Experiential Materials Mechanics Testing

Mechanical Test Of Different Types Specimens And Materials:

Metal:High-temperature alloys, Heat-resistant steel/Stainless steel, Titanium alloys, Aluminum alloys, Special metals, Welded joints

Polymers: Thermoplastics, Thermosetting plastics,Rubbers/Elastomers, Adhesives/Sealants

Composite materials:Fiber-reinforced plastics, Laminates / Hybrid composites etc.

Product Features

High Precision Load Cell

Testing load accuracy: class 0.5

Ensures stable and accurate test results

New designed Handheld Unit controller

320 x 240 LED display

Can display Load, displacement, speed. Can control test start, test stop, test zero clearing and so on

Easy and convenient operation

Germany NOVO Displacement sensor

The hysteresis telescopic non-contact displacement sensor has an effective range (maximum displacement) of ±75mm. When used with a control system, the indication error is ±0.5%FS. The sensor is calibrated before leaving the factory.

Upper And Lower Rods

Both the upper and lower rods are equipped with automatic centering ball seats, which can effectively ensure the coaxiality of the upper and lower chucks

Measurement And Control System

It has high-speed processing capability, low noise, high anti-interference, and high sampling frequency of 100 times/s.

The resolution of force and deformation measurement is high (24 bits), and the inside and outside are not divided into files, and the resolution does not change in the whole process. The user can set it according to the test requirements.

High Temperature Furnace

It has high-speed processing capability, low noise, high anti-interference, and high sampling frequency of 100 times/s.

The resolution of force and deformation measurement is high (24 bits), and the inside and outside are not divided into files, and the resolution does not change in the whole process. The user can set it according to the test requirements.

Creep testing machine Model Selection

What Can Our Creep Testing Machine Measure?

General FAQs

What is a creep testing machine?

A creep testing machine is specialized equipment designed to evaluate the long-term deformation behavior of materials under constant stress (or strain) and specific temperature conditions. It measures parameters like creep strain, creep rate, and rupture life to predict material performance in high-temperature or long-service environments, producing characteristic creep curves (strain vs. time) for analysis.​

They are widely used in industries where materials endure sustained stress and elevated temperatures, including:​

Aerospace: Testing turbine blades, spacecraft structural materials​

Energy/Power: Evaluating boiler components, pipelines, and steam turbines​

Automotive: Assessing engine parts and exhaust system materials​

Metallurgy/Composites: Researching creep behavior of metals, ceramics, plastics, and CMCs (Ceramic Matrix Composites)​

Petrochemical: Validating pressure vessel integrity under high-temperature conditions.

Electromechanical machines : Covering test durations up to 10,000 hours and loads up to 250 kN​

Multi-station machines: Enabling simultaneous multiple tests for efficiency​

High-temperature models: Operating at temperatures ranging from -80 °C to +2,000 °C for extreme environment testing

Ensure the machine is placed on a stable, level surface to avoid measurement errors.​

Inspect mechanical components (gears, screws, grips) for wear, and verify proper lubrication.​

Check calibration status of load cells, temperature sensors, and extensometers (critical for data accuracy).​

Confirm specimen alignment: Radial eccentricity must be ≤ 0.02–0.05 mm to prevent premature failure.

Use standardized geometries (round/rectangular) with uniform cross-sections and specified gauge lengths.​

Ensure proper surface finish (no defects) via machining/polishing to avoid stress concentrations.​

Match specimen condition (heat treatment, grain size) to real-world service status.​

Secure specimens with grips that prevent slippage without introducing extra stress

What test parameters need to be set?

Stress level: Constant load (tensile/compressive) based on material service conditions​

Temperature: Ranging from ambient to 2,000 °C, with uniformity ≤ 2–3 °C across the specimen gauge length.​

Test duration: Short-term (≤10,000 hours) or long-term (>10,000 hours) depending on application.​

Atmosphere: Air, inert gas, or vacuum (to prevent material oxidation at high temperatures).

Critical components (load cells, temperature sensors, extensometers): Annual calibration​

Full machine calibration: Every 2–3 years (or after major maintenance/repair)​

Verify calibration before high-precision tests or if data anomalies occur

Check thermostat settings and recalibrate temperature sensors.​

Inspect heating elements for aging/damage and replace if necessary.​

Tighten loose electrical connections and ensure stable power supply.​

Verify thermocouple placement (near the specimen) and avoid radiation interference.

For hydraulic systems: Check oil level/quality, eliminate leaks, and inspect pumps/cylinders.​

For mechanical systems: Lubricate worn gears/screws or replace damaged components.​

Recalibrate loading sensors to correct force measurement inaccuracies.

Loose/broken data cables: Reconnect or replace damaged cables.​

Faulty sensors: Inspect extensometers/temperature sensors and replace if malfunctioning.​

Software issues: Verify settings, update, or reinstall data acquisition software.​

Computer system errors: Check hardware/OS stability and restart if needed.